4
11..
UPON
POWER-UP the unit enters “Measure Mode” displaying applied pressure.
22..
TW
O FUNCTION is available to the user in “Measure Mode”.
A. Zero Adjustment Mode: In the measurement mode,
the pressure connection is open to the atmosphere and
$
key is
pressed for more than 3 seconds in order to shift to zero adjustment
mode for zero point adjustment of the differential pressure sensor
If the zero point adjustment is performed correctly the message “ADJ”
will be displayed for 2 seconds, and the display will return to the meas-
urement mode.
B. Key Lock
Function Key Manual Indicator
Setting of key lock MODE+
$
one second LoC (Key invalidity)
Release of key lock MODE+
$
one second UnL (Key invalidity)
Operation during keylock
Function Key Manual Indicator
Zero adjust. mode
$
key greater than 3 sec LoC (Key invalidity)
Hold value reset
#
key greater than 3 sec LoC (Key invalidity)
33..
FI
VE FUNCTIONS available to the User via “Setting Mode”. To enter the “Set-
ting Mode” hold (M) key for more than 3 seconds.
(See last page for complete Setting Mode menu.)
A. Filter (Damping)
The filter is based on the moving average of the pressure data to de-
crease display “bounce” and to smooth the analog output due to system
pressure fluctuations at the user’s discretion.
Five selections: (0, 2, 4, 8 and 16 seconds), Use
$#
keys to change value.
If “0” is selected the filter is not applied.
B. Re-scaling in “inH
2
O” units: “Pressure Display Mode” allows for zero
(4mA) and span (20ma) adjustment of –10 to +110% Span respectively.
Note: 1. See menu schematic on last page for detail.
2. Must be in “Pressure Display Mode” option within “Setting Mode,
this is noted on the screen by
Use
$#
keys to move between “Pressure Display Mode” and
“Linear Display Mode” which is for re-scaling inArbitrary” units.
Note: Holding the button for more than
3 seconds returns to display mode.
n non
-